Page 5 AD-4408C Weighing Indicator
3. Specifications
3.1. Analog Section
Input sensitivity 0.15 μV/d or greater
Input voltage range -35 mV to +35 mV (-7 mV/V to +7 mV/V)
Zero adjustment range -35 mV to +35 mV (-7 mV/V to +7 mV/V)
Load cell excitation voltage 5VDC 5%, 120 mA with remote sensing capability
Load cell drive capacity Maximum 8 x 350 Ω load cells
Temperature Zero 0.02 μV /°C typ., 0.1 μV /°C max.
coefficient Span 3ppm /°C typ., 15ppm /°C max.
Non-Linearity 0.005 % of full scale
A/D conversion method Delta-sigma method
A/D resolution count Approximately 16,000,000 counts
Display resolution 999,999 d max.
Sampling rate 100 times/second
3.2. Digital Section
Display element
• Measurement display: 6-digit 7-segment green LED
• Character height: 14.6 mm
• Polarity display: 1 green LED
• Status indicator: 6 red LEDs
Measurement display • Switches between NET and GROSS
• Selectable decimal places (10
1
, 10
2
, 10
3
, 10
4
, 10
5
)
• Overflow display
All the digits turn OFF. (When the polarity is negative, the
minus sign appears at the highest-order digit.)
Status indicators ZERO, STABLE, GROSS, NET, HOLD,
Key switches ZERO, TARE, NET/GROSS, F, ON/OFF, ENTER, CAL