
Page 35 AD-4408C Weighing Indicator
9. Interface
9.1. Standard Serial Output
The standard serial output with 0 to 20 mA can be connected to a device such as an
A&D printer or remote display. The interface has no power source. Therefore, the
external devices to be connected must have a power source. The setting values of the
standard serial output are changed in the general functions Cl f01 to Cl f03.
Type 0 to 20 mA current loop
Data bits 7 bits
Start bit 1 bit
Parity 1 bit, even
Stop bit 1 bit
Baud rate 600, 2400 bps
Code ASCII
9.1.1. Connection
Fig.7 Standard serial output pin connection
Fig.8 Standard serial output internal circuit
NOTE: The standard serial output connection has no polarity.
When a shielded cable is used, connect the shield to the F.G. terminal.
Standard
serial out
ut
Frame ground
C.L.
F.G.
8 9
10
Internal circuit
C.LOOP
output
terminal
No polarity
Standard
serial out
ut
Komentarze do niniejszej Instrukcji