AD-4408C Weighing Indicator Page 8
4. Description of Each Part
4.1. Front Panel
Fig.1 Front panel
Displays
No. Name Description
1 Main display Displays the weight value, setting values, error messages, etc.
2 ZERO Turns ON when the weight value is in the center of zero.
3 STABLE
Turns ON when the weight value is stable.
The stability conditions can be changed in the calibration-related
functions.
4 GROSS Turns ON when the displayed value is a gross weight.
5 NET Turns ON when the displayed value is a net weight.
6 HOLD
Turns ON when the weight value is being held.
Two hold modes can be selected in the general functions, either
normal hold or peak hold.
7
The function for this indicator can be selected in the general functions,
depending on the purpose.
8 UNIT
Place a unit label.
A unit used for weighing is set in the calibration-related functions.
9 Capacity plate Place a weighing capacity plate.
10 CAL
The key to enter the calibration mode.
The key is concealed by a sealable cover.
To avoid unintentional operations, keep the cover attached
except when entering the calibration mode.