Page 49 AD-4408C Weighing Indicator
Error status flag
When an error occurs to the AD-4408C, the remote READY flag (RX007B) becomes
inactive and the error status flag (RX007A) becomes active to inform the master station
of the error occurrence.
The master station requests to reset the error status flag, by activating the request flag
of error reset (RY007A).
Fig.14 Resetting the error status flag
Remote READY flag (RX007B)
Error status flag (RX007A)
Error reset request (RY007
Error code (RWr0006)
Error code (RWr0006) and
Error sub-code (RWr0007)
will be cleared to 0.
Error sub-code (RWr0007)
Re
q
uest fla
g
of error reset
(
RY007A
)