Page 61 AD-4408C Weighing Indicator
Function No.
Parameter or
Setting range
Description
Default
value
User
setting
Calf13
1 to 3
Condition to judge over when the gross
weight is negative.
A/D negative over or
1: Gross weight < -999999
2: Gross weight < negative weighing capacity
3: Gross weight < -19 d
1
Calf14
1 to 2
Condition to judge over when the net weight
is negative.
A/D negative over or
1: Net weight < -999999
2: Net weight < negative weighing capacity
1
Calf15
0 to 1
Select whether or not to clear the zero
value.
0: Disables.
1: Enables.
1
Calf16
0 to 1
Select whether or not to perform zero upon
power-ON.
0: Disables.
1: Enables.
0
Calf17
-7.00000 to
7.00000
Input voltage from a load cell at zero.
X.XXXXX mV/V
0.00000
Calf18
0.00001 to
9.99999
Input voltage from a load cell at span
X.XXXXX mV/V
2.00000
Calf19
1 to 999999
The value of the calibration weight
corresponding to the span input voltage
20000
Calf20
0 to 1
Select whether or not to output when the
gross weight is negative.
0: Enables output.
1: Disables output.
0
Calf21
0
to 1
Select whether or not to restrict
network-related communication.
0: Disables restriction.
1: Enables restriction.
0
Calf22
1 to 2
Second header for serial output
1: GS / NT / TR
2: G_ / N_ / T_ (_ space)
1
Przeglądanie stron 62