Page 25 AD-4408C Weighing Indicator
Function No.
Parameter or
Setting range
Function Description
Default
value
Calf06
0.0 to 5.0
Zero
tracking
time
Performs zero tracking using this setting in
combination with the setting of the zero
tracking band.
When 0.0, zero tracking will not be
performed.
Unit: second in 0.1 increments
0.0
Calf07
0.0 to 9.9
Zero
tracking
band
Performs zero tracking using this setting in
combination with the setting of the zero
tracking time.
When 0.0, zero tracking will not be
performed.
Unit: d (minimum division) in 0.1 increments
0.0