66
DATA TRANSFER MODE
There are five kinds of data transfer modes for RS-232C/422/485; “stream,” “auto print,” “manual print,”,“command”, and
“output per sampling.”
[Stream], [Auto Print], and [Manual Print]
The functions of these transfer modes are the same as described in 8-3 Standard Serial Output.
[Command]
This mode analyzes the “command” received from the host computer, etc. provides relevant processing, and responds
with a result. For example, if an “MT” command is received, the tare will be subtracted and its result (whether tare
subtraction is allowed or not) will be sent as a response.
The address capability is available only in this mode.
[Output per Sampling]
There are two types; gross and net. This mode outputs the setpoint state as well as weights. This is stream operations
outputting at a rate of 100 times/second in synchronization with sampling.
TRANSMISSION FORMAT
There are three kinds of transmission format; “A & D standard format”, “Accumulation data format”, and
“Output-per-Sampling format.” ( Outputs data 100 times per second synchronized with sampling.)
The A & D standard format used for this weighing indicator consists of two headers, data, unit, and a terminator.
A & D Standard Format Example
Header-
1
Header-
2
8 Digits Including Data, Polarity, and
Decimal Point
Unit Terminator
Gross S T , G S , + 0 0 1 2 3 4 5 k g CR LF
Header-2 is
"GS"
Net S T , N T , + 0 0 1 0 0 0 0 k g CR LF
Header-2 is
"NT"
Tare S T , T R , + 0 0 0 2 3 4 5 k g CR LF
Header-2 is
"TR"
With
decimal
point
S T , G S , + 0 1 2 3 . 4 5 k g CR LF
"." in
numerical
data
Plus
overflow
O L , G S , + SP SP SP SP . SP SP k g CR LF
Header-1 is
"OL"
Minus
overflow
O L , G S , - SP SP SP SP . SP SP k g CR LF
Header-1 is
"OL" and
polarity is
"-"
Unstable U S , G S , + 0 1 2 3 . 4 5 k g CR LF
Header-1 is
"US"
Output OFF
data
O L , G S , + SP SP SP SP . SP SP k g CR LF
Same as
plus
overflow
Komentarze do niniejszej Instrukcji