Strona 1 - PT-Series

Pipette Accuracy TesterPT-Series AD-4212B-PT AD-4212A-PT FX-300i-PT INSTRUCTION MANUAL 1WMPD4001794D

Strona 2

82-2 FX-300i-PT 1. Place the balance on a solid table. For details on the installation site, refer to “3. PRECAUTIONS BEFORE USE”. .

Strona 3 - CONTENTS

9 4. Install the lower part of the evaporation trap on the weighing unit, mating the tabs with the holes on the base. 5. Attach the upper pa

Strona 4 - 1. INTRODUCTION

103. PRECAUTIONS BEFORE USE To get the optimum performance from the balance and acquire accurate weighing data, note the following, especially when

Strona 5 - FX-300i-PT

114. ANTI-STATIC TREATMENT An anti-static treatment has been applied to the evaporation trap. Cleaning it with alcohol will reduce the effect of the

Strona 6 - Main Window

126. CONNECTING A PC TO THE BALANCE To make a connection between a PC and the balance, use the RS-232C cable or use both the RS-232C cable and the

Strona 7

136-2 Checking COM Port Numbers COM port numbers vary depending on the PC used. Use the Windows Device Manager to check COM port numbers. NOTE W

Strona 8 - 2. ASSEMBLING THE TESTER

147. CALIBRATING THE BALANCE For details on calibration, read the balance instruction manual in addition to this manual. Before calibration, refer

Strona 9

157-2 FX-300i-PT 1. Remove the 30 mL sample cup and leave only the sample cup holder (30 mL) inside the evaporation trap as shown to the right.

Strona 10 - 2-2 FX-300i-PT

168. PREPARATION BEFORE USE The measurement environment recommended by ISO 8655 is as follows. Temperature: Constant between 15°C and 30°C, changes

Strona 11

178-4 Using the Thermometer Each part name Material: Main body, battery cover: Acrylonitrile-butadiene- styrene (ABS) Button: Rubbe

Strona 12 - 3. PRECAUTIONS BEFORE USE

© 2015 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into

Strona 13 - 5. INSTALLING WinCT-Pipette

18Replacing The Battery The battery is not installed when you purchase the thermometer. Properly install the battery in the thermometer by the follow

Strona 14 - 6-1 Connecting a PC

19Measuring the temperature 1. Press the [ON/OFF] button to turn the power on. 2. Remove the sheath holder when it is attached to the sheath. 3. M

Strona 15

20When measuring the gas Place the sensor in the location where you would like to measure the temperature and wait until the display becomes stable.

Strona 16 - 7. CALIBRATING THE BALANCE

21 The thermometer automatically displays the current measurement temperature if no operation is made for 30 seconds while on the maximum or minimum

Strona 17 - 7-2 FX-300i-PT

22How to use the sheath holder Use the sheath holder to protect the sheath when you carry the thermometer. Be sure the sheath holder is removed from

Strona 18 - 8. PREPARATION BEFORE USE

239. MEASUREMENT 9-1 Measuring Procedure Using the predetermined pipette volume, perform the measurement the specified times. ISO 8655 recommends ten

Strona 19 - 8-4 Using the Thermometer

24 To reduce errors caused by water temperature Use the correct Z factor. To obtain the correct Z factor, measure the water temperature using the ac

Strona 20

2510. AD-4212B-PT / AD-4212A-PT The AD-4212B-PT / AD-4212A-PT is designed exclusively as a pipette accuracy tester. Although the performance as a

Strona 21

2610-2 Display Lock Function To use this function, set “Display lock function (ploc)” to “ON (1)”. Why use this function This function is used to co

Strona 22

2711. TROUBLESHOOTING  Displayed values are unstable. The AD-4212B-PT / AD-4212A-PT has a minimum weighing value of 0.001 mg / 0.1 mg and is influe

Strona 23

1CONTENTS 1. INTRODUCTION ...

Strona 24 - Specifications

2812. SPECIFICATIONS AD-4212B-PT AD-4212A-PT FX-300i-PT Weighing capacity *1 110 g / 31 g / 5.1 g *2 110 g 320 g Minimum weighing value 0.1 mg / 0

Strona 25 - 9. MEASUREMENT

2912-1 Options Name and components Model 5-mL sample cup (with a water-absorbent sheet installed) 5-mL sample cup 5 pcs. Lid for 5-mL sample cup

Strona 26 - Operator

3013. APPENDIX The table below lists the relation between the ISO 8655 requirements and the PT series pipette accuracy testers. ISO 8655 requiremen

Strona 30 - 12. SPECIFICATIONS

3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013, JAPAN Telephone: [81] (3) 5391-6132 Fax: [81] (3) 5391-6148 A&D ENGINEERING, INC

Strona 31 - 12-1 Options

21. INTRODUCTION The PT series Pipette Accuracy Tester performs an accuracy test of the volume of a pipette in accordance with ISO 8655. This manual

Strona 32 - 13. APPENDIX

3FX-300i-PT 1-2 Description of Each Component  Evaporation trap A chamber to prevent the evaporation of the dispensed test liquid

Strona 33

41-3 WinCT-Pipette, Software for Pipette Accuracy Testing Features The mass value of the distilled water dispensed from a pipette is transmitted f

Strona 34

51-4 Function of the Evaporation Trap In the pipette accuracy test using the gravimetric method, to obtain the mass of the test liquid, distilled wa

Strona 35

6Stand Angle adjustment knob Weighing unitDisplay Connectioncable Sample cup holder (5 mL)5 mL sample cup 30 mL sample cup Sample cup holder (30 mL

Strona 36

7 4. Install the lower part of the evaporation trap on the weighing unit, mating the tabs with the holes on the base. 5. Attach the upper pa