AD-4430B - 27 -
Function No.
Parameter
and range
Function name Description
Default
value
CAL16
0 to 1
Zero setting when
power is turned on
Select whether or not to perform zero setting when
power is turned on.
0 : Disables.
1 : Enables.
0
CAL17
-7.0000
to 7.0000
Input voltage at zero
Input voltage from a load cell at zero. Unit : mV/V.
This value is determined in zero calibration during
the calibration with an actual load.
0.0000
CAL18
0.0001
to 9.9999
Input voltage at span
Input voltage from a load cell at span. Unit : mV/V.
This value and the value of CAL19 are determined
in span calibration during the calibration with an
actual load.
3.2000
CAL19
1 to 99999
Calibration weight
value corresponding
to Input voltage at
span
The calibration weight value corresponding to the
Input voltage at span of CAL18. When performing
digital span, CAL17, CAL18 and CAL19 are
required. The decimal point position is the same as
the setting of CAL02.
32000
NOTE:
*1 Record the setting values of CAL17, CAL18 and CAL19 in the “Setting List” at the end of the
manual to prepare against a failure.
*2 By changing the parameters of CAL17, CAL18 and CAL19, “zero calibration” and “span
calibration” can be adjusted optionally. (Digital span accuracy approximately 1/5000. The
accuracy varies depending on the load cell output accuracy and the conditions of calibration.)
Except for an emergency, perform calibration with an actual load.
CAL26
9.7500
to 9.8500
Gravity acceleration
of place of calibration
Gravity acceleration of the place of calibration.
Unit : in m/s
2
.
9.8000
CAL27
9.7500
to 9.8500
Gravity acceleration
of place of use
Gravity acceleration of the place where the scale is
being used. Unit : in m/s
2
.
9.8000
CAL28
0 to 1
Disable averaging
hold
0 : Disables.
1 : Enables.
0
CAL19
CAL18
CAL17
Input voltag
Displayed weigh
0