AD-4430B - 23 -
5.3.5. The Actual Load Linearization Function
Set the digital linearization by loading/unloading masses.
Warm up the module for at least ten minutes to avoid the effects of temperature drift.
The input order should proceed from the smallest mass to the largest mass.
Step 1 Press the + key to enter to the calibration mode and
display . Press the key to start the calibration
and display . Select using the key
and press the key.
Step 2 is displayed.
If monitoring the current weighing value, press the key.
When pressing the key again, is display.
Step 3 Placed nothing on the pan and wait for the stabilization ( LED ).
Press the key. is displayed for approximately
two seconds.
Step 4 is displayed.
If you want to check the current weighing value, press the key.
When pressing the key again, is displayed.
Press the key. The weight value (the current weighing
capacity) is displayed and the least digit of the value blinks.
Correct the value using the and key so as to be the
weight value used.
Step 5 Place the weight on the pan. Wait for the stabilization ( LED ).
Press the key. is displayed for approximately
two seconds.
Step 5 is displayed. Repeat step 4 and step 5. The procedure
proceeds in order of → → .
Step 7 If finishing the input, proceed to step 8.
If you want to re-input the digital linearization, select the input point
using the key. All data following the new input point will be
cleared.
Step 8 Press the key. is displayed and the inputted
data will be stored in the FRAM. At the same time, the calibrated
data is also refreshed. Press again the key to return to
weighing mode.
If
X
is displayed, an error has occurred. X : number.
Refer to “5.3.8. Error Codes For The Calibration” to take some measures.
The blinking decimal point means that the current value is not the weight value.
Example