Przeglądanie stron 102
AD-4402 Page 101
Prucedure
Caution
Do not intialize them while any operation. Cut off the power supply of other
system. When initializing the indicator, the output may change.
When intializing the indicator, do not turn off it before it is resetted.
To enter intialization
Step 1 Press and hold the ENTER key and press the key to display the
menu in a weighing mode.
Step 2 Select the menu "Initialization" using the key, ENTER key and
ESC key.
Category address: [Maintenance] - [Initialization]
To enter intialization
Step 1 Select the menu intialization using the key, ENTER key and
ESC key.
Menu: RAM / Code / Function / CAL / All
Step 2 Select "yes" to initialize them using the key, ENTER key.
Step 3 Wait for the indicator resetted.