A&D GR Series Instruction Manual
GB 30
8. If you add a few more items, the ACAI indicator turns on. (The ACAI
indicator turns off if in overload.)
9. The balance re-calculates the unit weight while the ACAI indicator is
blinking. Wait and do not touch the items on the pan until the ACAi
indicator turns off automatically.
10. You are now able to count items with a more accurate unit weight.
11. If you add a few more items, proceed to step 8. The balance
recalculates a more accurate unit weight.
ACAI indicator
Detecting items
Calculating unit weight
Counting mode
6.3. Percent Mode (%)
Selecting the unit of percent mode
1. Select the unit % using the [MODE] key. If the percent mode can not be selected, refer to
Section “5. Weighing Units”. (% : percent)
Storing 100% weight
2. Press the [RANGE] key to enter the sampling mode.
3. Place a container on the weighing pan, if necessary. Press [RE-ZERO] key to cancel the
container weight and possible errors. The balance displays 100 0 %.
4. Place the item of 100% on the weighing pan or in the container.
5. Press the [PRINT] key to store this 100% weight.
6. Remove the item from the pan.
Reading percentage
7. You are now able to read the percentage based on the stored 100% weight.