7. Maintenance
7-1 Cleaning
Caution
Before cleaning, turn off the power and remove the AC adapter.
Do not pour water on the instrument or use water for cleaning. This instrument is not
waterproof.
Do not sterilize the instrument by gaseous sterilization or using an autoclave.
Do not use solvents such as thinner and benzine for cleaning. That may deform or
discolor the instrument.
Clean the instrument as described below:
Clean the instrument exterior with a dry soft cloth.
If the instrument is stained with blood, chemical substance, or dirt, wipe the instrument off using a cloth
dampened with a mild detergent.
For cleaning a patient's body site to apply a sensor, follow the procedure of each hospital.
To disinfect the cuff, wipe the surface which comes into contact with the skin with alcohol such as 70%
isopropyl. Do not wash the cuff.
Discard the disposables after use. To disinfect the optional SpO
2
sensor or the extension cable, use
alcohol such as 70% isopropyl or bleach diluted by 10%.
Wipe the ECG extension cable and ECG patient cable with alcohol such as 70% isopropyl.